1. Collect 1 mL blood into each of the 4 separate
QuantiFERON collection tubes. As 1 mL tubes draw
relatively slowly, keep the tube on the needle for 2-3
seconds once the tube appears to have completed filling
to ensure that the correct volume is drawn.
blood can be collected using a syringe and 1 mL blood
transferred to each of the 4 tubes.
if a “butterfly” needle is being used to collect the
sample, a “purge” tube should be used to ensure that the
tubing is filled with blood prior to the QuantiFERON
tubes being used.
2. Mix the tube by shaking 10 times just firmly enough to
coat the entire inner surface of the tubes.
3. Samples must be maintained at room temperature and
received in the laboratory within 16 hours of
collection. After receipt in the laboratory, samples
are incubated at 37 degrees Celsius for one day prior to
interferon assay. STAT testing is not available.
Critical Room Temperature (CRT).