Nickel, Serum

Test Code: TD_4271
CPT Code(s): 83885

$49

Test Details

Alternative Name(s)

N/A

Test Includes

Nickel

Methodology

Inductively coupled plasma-mass spectrometry (ICP/MS)

Rejection Criteria

Samples collected in serum separator tube
Serum not separated from cells within 2 hours of collection
Samples collected in plain red top tube
Heparin (green top) plasma Samples submitted in non-trace metal-free tubes

Reference Range(s)

<=10.0 UG/L

Specimen Requirements

Preferred Specimen

Trace metal-free (royal blue top) tube with no additive

Minimum Volume

0.5 mL serum

Sample Stability

If sample is collected and stored in the appropriate container, trace element values do not change with time

Transport Temperature

Room temperature

Collection Instructions

2 mL serum from a royal blue trace metal-free tube without
additive. Allow tube to clot for at least 60 minutes, then
centrifuge sample and separate serum from cells within
2 hours of collection. Transfer 2 mL serum to a plastic
transport tube. Clearly label tube as serum from a royal
blue top tube. Room temperature.

Diet, medication, and nutritional supplements may introduce
interfering substances. Patient should be encouraged to
discontinue nutritional supplements, vitamins, minerals,
and non-essential over-the-counter medications (upon the
advice of their physician).

Alternative Specimen

N/A